平成16年 7月26日作成

医療用具改修の概要

(クラス III)



  1. 一般名及び販売名

    一般的名称:超電導式磁気共鳴画像診断装置
    販売名  :下記(1)から(6)までの東芝MRI
    (1)東芝MRI MRT-50A
    (2)東芝MRI FLEXART  MRT-50GP
    (3)東芝MRI VISART   MRT-200シリーズ
    (4)東芝MRI EXCELART MRT-2000
    (5)東芝MRI EXCELART MRT-2001
    (6)東芝MRI EXCELART MRT-1000
    
    
  2. 対象ロット、数量及び出荷時期

    型式名    :MRT-50A
    Ser No.    :0522172,0532182,0622265,1512275,1522286,1562309,2532026,2582037
                   3592046,3597007,7562010,9532090,9592135
    型式名    :MRT-50GP
    Ser No.    :0552035,0572036,0582037,0622038,1512006,3572062,4562010,4592020
                   6602012,6627002,7512001,7622004,8612014,9512015,9517002,9612029
                   9622001
    型式名    :MRT-200
    Ser No.    :0552034,0572040,0592043,0602013,0622014,0627054,1512015,1562068
                   1592002,1592003,2522014,3542047,3572050,4532063,4532065,4572066
                   4592067,4612069,5522001,5532005,5532006,5542009,5582011,5592012
                   5592016,5612017,5612018,6522025,6557018,6572034,6582035,6582037
                   6592039,6602043,7512002,7512053,7522055,7522059,7532062,7532065
                   7532067,7542070,7542072,7562082,7572084,7572086,7612095,7622093
                   7622096,7622099,8512101,8522001,8522103,8522105,8522106,8522109
                   8532002,8562119,8572123,8582001,8582124,8592004,8602005,8612006
                   8612128,8622008,9522009,9522011,9522012,9542004,9542006,9547048
                   9562009,9592006,9602007
    型式名    :MRT-2000
    Ser No.    :0512007,0512008,0512010,0522011,0522012,0532004,0532013,0542015
                   0552016,0552017,0572019,0582023,0582024,0592026,0592027,0602030
                   0612032,0612034,0622035,0622036,1512037,1522039,1522040,1522041
                   1532042,1532043,9522001,9532002,9532004,9532005,9572006,9572007
                   9572008,9582009,9582010,9592011,9592013,9592014,9592016,9612001
                   9612002,9612005,9622003,9622006
    型式名    :MRT-2001
    Ser No.    :0532001,0562002,0597001,0597002,0602004,0622006,0622007,1512001
                   1522002,1532004,1532005,1542009,1552010,1562012,1572009,1572014
                   1572015,1582010,1592012,1592013,1592014,1592015,1592016,1592018
                   1592019,1602020,1612022,1612023,1612024,1612026,1612027,1622028
                   1622029,2512019,2512020,2512021,2512030,2512031,2512032,2512033
                   2522023,2522034,2522035,2522037,2522038,2532024,2532025,2532026
                   2532039,2532040,2532041,2562043,2572001,2572027,2572044,2582001
                   2582045,2592002,2592028,2592029,2592046,2592048,2602002,2602003
                   2602030,2612004,2612049,2612050,2612051,2612052,2622005,2622031
                   2622032,2622033,2622053,2622054,3512034,3512035,3512036,3512055
                   3522037,3522056,3522057,3522058,3532003,3532059,3542038,3542060
                   3542061,3542062,3552063,3562006,3562007,3562039,3562064,3572008
                   3572065,3572066,3572067,3582009,3582068,3592040,3592041,3592069
                   3602070,3602071,3612072,4522073
    型式名    :MRT-1000
    Ser No.    :1532001,1532002,1542004,1542005,1552006,1562007,1572008,1572009
                   1582010,1582012,2512017,2512019,2522021,2522022,2532027,2532028
                   2542030,2572032,2582033,2582034,2582035,2592036,2622037,2622038
                   3512039,3522040,3522041,3522042,3532043,3532045,3532047,3532048
                   3532049,3532050,3542051,3542052,3552053,3552054,3562055,3572056
                   3572057,3572058,3592060,3592061,3592062,3612064,3612065,4512066
                   4532067
    数量       :306台
    出荷時期    :昭和62年06月〜平成15年11月
    
    
  3. 製造業者等名称

    会社の名称 :東芝メディカルシステムズ株式会社
    本社所在地 :栃木県大田原市下石上1385番地
    薬事の業態 :医療用具製造業
    製造所      :東芝メディカルシステムズ株式会社
    製造所の住所:栃木県大田原市下石上1385番地
    業許可番号 :09BZ6008
    
    
  4. 改修理由

    画像サーバー等(以下、サーバー)への情報の送信(DICOM規格)において、
    ソフトウェアの不具合により以下の条件が重なった場合に、患者IDの下2桁が
    落ちて送信されるという現象が起こることが判明しました。
    MRI装置側での患者ID表示および保存に問題は生じません。
    (1)患者IDの桁数を5桁または6桁で運用している。
    (2)患者さんの誕生日が1月1日である。
    (3)装置内部での患者名のデータ長が10、12、14、34、36または38バイトである。
    このため当社としては、すべての該当装置に改善したソフトウェアのインストールを
    改修として実施することとしました。
    
    
  5. 危惧される具体的な健康被害

    前述のように3つの条件が全て重なった場合に発生する稀な事象です。患者IDの下2桁が
    落ちて送信されることによりサーバーでの患者IDによる画像検索ができなくなる
    場合が考えられます。その場合は通常は氏名検索により画像を検索することが可能です。
    またMRI装置内部には画像が保存されております。
    このため健康被害への影響は無いと考えています。
    なお、本件による健康被害の報告は受けておりません。
    
    
  6. 改修開始年月日

    平成16年 07月 24日 (情報提供の開始)
    
    
  7. 効能・効果又は用途等

    磁気共鳴画像診断に用いる装置で、医師もしくは検査技師等有資格者が使用します。
    
    
  8. その他

    納入しました医療機関は全て特定しており、改善したソフトウェアのインストールを、
    改修として実施させていただきます。
    
    