一般名及び販売名
一般的名称:超電導式磁気共鳴画像診断装置 販売名 :下記(1)から(6)までの東芝MRI (1)東芝MRI MRT-50A (2)東芝MRI FLEXART MRT-50GP (3)東芝MRI VISART MRT-200シリーズ (4)東芝MRI EXCELART MRT-2000 (5)東芝MRI EXCELART MRT-2001 (6)東芝MRI EXCELART MRT-1000
対象ロット、数量及び出荷時期
型式名 :MRT-50A Ser No. :0522172,0532182,0622265,1512275,1522286,1562309,2532026,2582037 3592046,3597007,7562010,9532090,9592135 型式名 :MRT-50GP Ser No. :0552035,0572036,0582037,0622038,1512006,3572062,4562010,4592020 6602012,6627002,7512001,7622004,8612014,9512015,9517002,9612029 9622001 型式名 :MRT-200 Ser No. :0552034,0572040,0592043,0602013,0622014,0627054,1512015,1562068 1592002,1592003,2522014,3542047,3572050,4532063,4532065,4572066 4592067,4612069,5522001,5532005,5532006,5542009,5582011,5592012 5592016,5612017,5612018,6522025,6557018,6572034,6582035,6582037 6592039,6602043,7512002,7512053,7522055,7522059,7532062,7532065 7532067,7542070,7542072,7562082,7572084,7572086,7612095,7622093 7622096,7622099,8512101,8522001,8522103,8522105,8522106,8522109 8532002,8562119,8572123,8582001,8582124,8592004,8602005,8612006 8612128,8622008,9522009,9522011,9522012,9542004,9542006,9547048 9562009,9592006,9602007 型式名 :MRT-2000 Ser No. :0512007,0512008,0512010,0522011,0522012,0532004,0532013,0542015 0552016,0552017,0572019,0582023,0582024,0592026,0592027,0602030 0612032,0612034,0622035,0622036,1512037,1522039,1522040,1522041 1532042,1532043,9522001,9532002,9532004,9532005,9572006,9572007 9572008,9582009,9582010,9592011,9592013,9592014,9592016,9612001 9612002,9612005,9622003,9622006 型式名 :MRT-2001 Ser No. :0532001,0562002,0597001,0597002,0602004,0622006,0622007,1512001 1522002,1532004,1532005,1542009,1552010,1562012,1572009,1572014 1572015,1582010,1592012,1592013,1592014,1592015,1592016,1592018 1592019,1602020,1612022,1612023,1612024,1612026,1612027,1622028 1622029,2512019,2512020,2512021,2512030,2512031,2512032,2512033 2522023,2522034,2522035,2522037,2522038,2532024,2532025,2532026 2532039,2532040,2532041,2562043,2572001,2572027,2572044,2582001 2582045,2592002,2592028,2592029,2592046,2592048,2602002,2602003 2602030,2612004,2612049,2612050,2612051,2612052,2622005,2622031 2622032,2622033,2622053,2622054,3512034,3512035,3512036,3512055 3522037,3522056,3522057,3522058,3532003,3532059,3542038,3542060 3542061,3542062,3552063,3562006,3562007,3562039,3562064,3572008 3572065,3572066,3572067,3582009,3582068,3592040,3592041,3592069 3602070,3602071,3612072,4522073 型式名 :MRT-1000 Ser No. :1532001,1532002,1542004,1542005,1552006,1562007,1572008,1572009 1582010,1582012,2512017,2512019,2522021,2522022,2532027,2532028 2542030,2572032,2582033,2582034,2582035,2592036,2622037,2622038 3512039,3522040,3522041,3522042,3532043,3532045,3532047,3532048 3532049,3532050,3542051,3542052,3552053,3552054,3562055,3572056 3572057,3572058,3592060,3592061,3592062,3612064,3612065,4512066 4532067 数量 :306台 出荷時期 :昭和62年06月〜平成15年11月
製造業者等名称
会社の名称 :東芝メディカルシステムズ株式会社 本社所在地 :栃木県大田原市下石上1385番地 薬事の業態 :医療用具製造業 製造所 :東芝メディカルシステムズ株式会社 製造所の住所:栃木県大田原市下石上1385番地 業許可番号 :09BZ6008
改修理由
画像サーバー等(以下、サーバー)への情報の送信(DICOM規格)において、 ソフトウェアの不具合により以下の条件が重なった場合に、患者IDの下2桁が 落ちて送信されるという現象が起こることが判明しました。 MRI装置側での患者ID表示および保存に問題は生じません。 (1)患者IDの桁数を5桁または6桁で運用している。 (2)患者さんの誕生日が1月1日である。 (3)装置内部での患者名のデータ長が10、12、14、34、36または38バイトである。 このため当社としては、すべての該当装置に改善したソフトウェアのインストールを 改修として実施することとしました。
危惧される具体的な健康被害
前述のように3つの条件が全て重なった場合に発生する稀な事象です。患者IDの下2桁が 落ちて送信されることによりサーバーでの患者IDによる画像検索ができなくなる 場合が考えられます。その場合は通常は氏名検索により画像を検索することが可能です。 またMRI装置内部には画像が保存されております。 このため健康被害への影響は無いと考えています。 なお、本件による健康被害の報告は受けておりません。
改修開始年月日
平成16年 07月 24日 (情報提供の開始)
効能・効果又は用途等
磁気共鳴画像診断に用いる装置で、医師もしくは検査技師等有資格者が使用します。
その他
納入しました医療機関は全て特定しており、改善したソフトウェアのインストールを、 改修として実施させていただきます。