一般名及び販売名
一般的名称 :汎用超音波画像診断装置 販売名 :超音波診断装置 APLIO SSA-700A,SSA-770A
対象ロット、数量及び出荷時期
形式名 :SSA-700A 計: 478台 出荷時期 :平成 14年 6月 〜 平成16年 6月 Ser No. 2562016,2562018,2562022,2562030,2572050,2572051,2572052,2572053 2572054,2572055,2572056,2572057,2582067,2582068,2582069,2582070 2582078,2582079,2582080,2582081,2582082,2582083,2582084,2582085 2582086,2582087,2582088,2592089,2592090,2592091,2592092,2592093 2592095,2592096,2592098,2592099,2592100,2592101,2592102,2592103 2592104,2592106,2592108,2592109,2592110,2592784,2602117,2602121 2602150,2612162,2612191,2612192,2612194,2612196,2612199,2612200 2612209,2622211,2622224,2622225,2622247,2622248,2622249,2622253 2622254,2622258,2622259,2622260,2622262,2622275,2622276,2622277 2622281,2622283,2622284,2622285,2622287,2622288,2622289,3512294 3512295,3512296,3512299,3512310,3512311,3512312,3512319,3512330 3512331,3512332,3512334,3512335,3512336,3522097,3522107,3522118 3522134,3522135,3522137,3522141,3522179,3522184,3522339,3522340 3522345,3522350,3522355,3522374,3522375,3522376,3522378,3522382 3522383,3522387,3522390,3522391,3522392,3522393,3522394,3522395 3522396,3522397,3522398,3522399,3532132,3532145,3532146,3532151 3532166,3532167,3532168,3532169,3532170,3532171,3532172,3532173 3532174,3532175,3532177,3532178,3532180,3532181,3532182,3532183 3532413,3532415,3532417,3532419,3532422,3532427,3532428,3532435 3532436,3532437,3532438,3532439,3532440,3532441,3532442,3532443 3532444,3532446,3532448,3532450,3532451,3532452,3532455,3532456 3532459,3532462,3532463,3532464,3532466,3532467,3532473,3532474 3532475,3532476,3532477,3532478,3542480,3542481,3542492,3542493 3542494,3542495,3542513,3542514,3542520,3542521,3542522,3542523 3542532,3542540,3552541,3552543,3552549,3552556,3552561,3552562 3552582,3552592,3552593,3572127,3572563,3572564,3572565,3572581 3572583,3572584,3572585,3572586,3572600,3572601,3572602,3572603 3572604,3572605,3572606,3572607,3572608,3572609,3572610,3572629 3572630,3572631,3572632,3572633,3572644,3572645,3572646,3572647 3572649,3572654,3572655,3582662,3582663,3582664,3582665,3582666 3582667,3582668,3582675,3582676,3582677,3582678,3582679,3582680 3582689,3582692,3582693,3582694,3582695,3582696,3582697,3582698 3582700,3582702,3582703,3582704,3582705,3582707,3582717,3582726 3582727,3582735,3582736,3582738,3582750,3582751,3582752,3582753 3582754,3592037,3592142,3592414,3592756,3592760,3592761,3592762 3592763,3592764,3592765,3592766,3592767,3592768,3592769,3592770 3592775,3592776,3592778,3592779,3592780,3592781,3592782,3592785 3592786,3602806,3602809,3602813,3602814,3602815,3602816,3602818 3602819,3602820,3602821,3602822,3602823,3602824,3602825,3602827 3612846,3612850,3612861,3612862,3612864,3612865,3612866,3612867 3612868,3612869,3612873,3612883,3612885,3622511,3622905,3622906 3622907,3622908,3622909,3622910,3622911,3622912,3622913,3622914 3622915,3622916,3622919,3622920,3622921,3622922,3622923,3622924 3622925,3622926,3622927,3622928,3622929,3622930,3622931,3622932 3622933,3622934,3622936,3622937,3622958,3622959,3622969,4512987 4512988,4512989,4512991,4512992,4512993,4512994,4513008,4523034 4523036,4523044,4523045,4523048,4523049,4523050,4523052,4523053 4523054,4523055,4523061,4523062,4523063,4523064,4523065,4523066 4523067,4523089,4523098,4523099,4523118,4532136,4532195,4532777 4532802,4532803,4533119,4533120,4533121,4533122,4533123,4533124 4533125,4533126,4533127,4533129,4533130,4533131,4533132,4533133 4533134,4533135,4533136,4533137,4533138,4533139,4533140,4533141 4533142,4533143,4533148,4533149,4533150,4533151,4533152,4533153 4533154,4533155,4533156,4533157,4533158,4533159,4533160,4533161 4533162,4533163,4533164,4533165,4533166,4533167,4533168,4533169 4533170,4533171,4533172,4533173,4533174,4533175,4533176,4533177 4543178,4543191,4543192,4543213,4543218,4543221,4543233,4543239 4543240,4552884,4552889,4553243,4553254,4553255,4553256,4553257 4553258,4553259,4553277,4553278,4553280,4563282,4563283,4563285 4563287,4563288,4563295,4563316,4563319,4563320 形式名 :SSA-770A 計: 237台 出荷時期 :平成 13年 7月 〜 平成16年 5月 SerNo. 1572001,1572002,1572003,1572004,1572005,1572006,1582007,1582008 1582011,1582016,1582017,1582019,1582020,1582021,1582022,1582023 1582025,1582026,1582027,1592028,1592033,1592036,1592037,1592038 1592039,1592040,1592047,1592048,1592049,1592050,1592051,1592052 1592054,1592055,1592056,1592058,1592059,1592060,1592062,1592063 1612074,1612075,1612076,1612077,1612079,1612081,1622091,1622118 1622119,1622120,1622123,1622124,1622127,2512073,2512129,2512131 2512132,2512136,2512146,2512147,2512148,2512149,2512150,2512151 2512168,2522169,2522170,2522171,2522179,2522180,2522181,2522184 2522185,2522186,2522187,2522188,2522189,2522192,2532193,2532194 2532195,2532196,2532197,2532202,2532203,2532205,2532206,2532212 2532213,2532214,2532215,2532215,2532216,2532218,2532219,2532219 2542220,2542231,2542237,2542239,2552257,2552263,2552264,2552275 2552278,2552281,2552283,2562301,2562303,2562310,2562318,2562329 2562333,2562334,2572335,2572336,2572337,2572358,2582367,2582377 2582378,2582387,2592201,2592389,2592391,2592392,2592393,2592394 2592397,2592399,2592400,2592401,2592403,2592404,2592405,2592409 2592410,2592411,2592412,2592416,2592417,2592418,2592419,2592423 2592424,2602451,2612467,2612475,2612479,3512516,3512526,3512527 3512530,3512554,3522190,3522578,3522587,3522588,3522596,3522598 3522604,3522605,3542607,3542609,3552661,3572706,3572707,3572711 3572722,3572733,3572734,3582744,3582747,3582752,3582755,3582756 3582757,3582759,3582760,3582761,3582762,3582764,3582765,3582766 3582767,3592183,3592332,3592392,3592608,3592610,3592688,3592690 3592780,3592791,3592792,3602797,3602798,3602802,3602815,3602822 3602832,3602839,3612851,3612852,3612853,3612880,3622631,3622910 3622913,3622914,3622918,3622919,3622920,3622921,3622923,3622924 3622925,3622928,3622931,3622932,3622933,3622949,3622951,4523036 4532172,4532175,4532813,4532814,4533053,4543062,4543070,4543079 4543100,4543101,4543104,4552046,4552899
製造業者等名称
会社の名称 :東芝メディカルシステムズ株式会社 本社所在地 :栃木県大田原市下石上1385番地 薬事の業態 :医療用具製造業 業許可番号 :09BZ6008
改修理由
本装置は、検査中の音響パワーを表示しており、TI(Thermal Index)は温度上昇を表す量で 生体の温度が1℃だけ上昇するのに必要な音響パワーに対する現在の音響パワーの比です。 この装置とコンベックス式電子スキャンプローブ PVT-375AT との組み合わせにおいて、 TIを表示するためのデータテーブルに誤りがあり、小さく表示がされる 場合があることが判明しました。このため、データテーブルを修正する 対策を改修として実施することにしました。
危惧される具体的な健康被害
TIは診断に使用するものではなく、音響パワーをモニタするための1つの参考情報です。 実際の検査においては、医師もしくは検査技師が画像を見て音響パワーを 診断可能な範囲内でできるだけ低く設定するため、TIを見て音響パワーを上げることはありません。 なお、音響パワーは、生体内での温度上昇が安全な範囲になるよう制御されています。 上記の理由から、健康被害への影響は無いと考えております。 また、本件による健康被害の報告は受けておりません。
改修開始年月日
平成16年 7月16日(情報提供・改修の開始)
効能・効果又は用途等
超音波画像診断装置は超音波を利用して人体を検査または診断する手段を 提供する装置で、医師もしくは検査技師等有資格者が使用します。
その他
納入しました医療機関は全て特定しており、全対象装置に修正されたデータテーブルを インストールする改修を実施します。