平成29年 6月 6日作成 平成29年 8月29日訂正(*)
一般名及び販売名
一般的名称: 汎用超音波画像診断装置 販売名 : 超音波診断装置 ARIETTA Prologue
対象ロット、数量及び出荷時期
204H4348,204H4349,204H4351,204H4352,204H4355,204H4356,204H4357,204H4359,204H4360,204H4362,204H4363,2 04H4364,204H4365,204H4366,204H4367,204H4368,204H4369,204H4370,204H4371,204H4372,204H4373,204H4374,20 4H4375,204H4376,204H4377,204H4379,204H4380,204H4381,204H4382,204H4407,204H4408,204H4409,204H4410,204 H4412,204H4413,204H4414,204H4415,204H4416,204H4417,204H4418,204H4419,204H4455,204H4456,204H4458,204H 4459,204H4460,204H4461,204H4463,204H4464,204H4465,204H4466,204H4467,204H4468,204H4469,204H4470,204H4 471,204H4472,204H4473,204H4474,204H4475,204H4476,204H4477,204H4478,204H4479,204H4480,204H4481,204H44 82,204H4483,204H4484,204H4485,204H4486,204H4488,204H4489,204H4492,204H4493,204H4494,204H4495,204H449 6,204H4497,204H4498,204H4499,204H4500,204H4503,204H4504,204H4505,204H4506,204J6072,204J6073,204J6074 ,204J6075,204J6076,204J6077,204J6078,204J6079,204J6080,204J6081,204J6082,204J6084,204J6085,204J6094, 204J6095,204J6096,204J6097,204J6098,204J6099,204J6100,204J6101,204J6105,204J6106,204J6109,204J6110,2 04P2997,204W2257,204W2258,204W2259,204W2260,204W2261,204Y1720,204Y1721,204Y1722,204Y1723,204Y1724,20 4Y1725,204Y1726,204Y1727,204Y1728,204Y1729,204Y1731,204Y1732,204Y1734,204Y1735,204Y1736,204Y1737,204 Y1738,204Y1739,204Y1740,204Y1741,204Y1743,204Y1744,204Y1745,204Y1746,204Y1748,204Y1750,204Y1751,204Y 1752,204Y1753,204Y1754,204Y1755,20509714,20509715,20509716,20509717,20509729,20509730,20509731,20509 732,20509733,20509734,20509735,20509736,20509744,20509745,20509746,20509747,20509748,20509750,205097 51,20509752,20509753,20509754,20509755,20509757,20509758,20535611,20535612,20535625,20535626,2053562 9,20535630,20535631,20535632,20535633,20535634,20535635,20535636,20535637,20535646,20535647,20535648 ,20535649,20535650,20570684,20570685,20570686,20570687,20570688,20570689,20570690,20570691,20570692, 20570693,20570698,20570699,20570700,20570701,20570702,20588651,20588652,20588653,20588654,20588655,2 0588656,20588657,20588658,20588659,20588660,20588661,20588662,20588663,20588664,20588665,20588666,20 588667,20588668,20588669,20588670,20588671,20588672,20588673,20588674,20588675,20588676,20588677,205 88678,20588679,20588680,20588681,20588682,20599605,20599606,20599607,20599610,20599611,20599612,2059 9613,20599614,20599615,20599616,205A1090,205A1091,205A1092,205A1093,205A1094,205A1095,205A1096,205A1 097,205A1098,205A1099,205A1100,205A1101,205A1102,205A1103,205A1104,205A1105,205A1106,205A1107,205A11 08,205A1109,205A1110,205A1111,205A1112,205A1113,205A1114,205A1115,205A1116,205A1117,205A1118,205A111 9,205A1120,205A1121,205A1122,205A1123,205A1156,205A1157,205A1158,205A1159,205A1160,205A1161,205A1170 ,205A1171,205A1172,205A1173,205A1174,205A1175,205A1178,205A1179,205A1180,205A1181,205A1183,205A1185 数量 :300台 出荷時期:平成28年3月〜平成29年4月
製造販売業者等名称
製造販売業者の名称 : 株式会社日立製作所 ヘルスケアビジネスユニット 製造販売業者の所在地: 東京都台東区東上野二丁目16番1号 許可の種類 : 第一種医療機器製造販売業 許可番号 : 13B1X90003
改修理由
一部の製品において、バックエンドユニット(画面側)とフロントエンドユニット(プローブを接続する側) 間の接続コネクタに接触不良(汚れ等による)が発生した場合、動作が不安定(予期せずフリーズしたり シャットダウンする、プローブを認識しない等)になることが判明したため、ソフトウェアの更新を実施いたし ます。
危惧される具体的な健康被害
現象が発生した場合、装置を使用できない状態になり診断の遅延につながる可能性がありますが、再起動する ことにより使用可能な状態となりますので、重篤な健康被害が発生する可能性はないものと考えます。 なお、これまでに本事象による健康被害は報告されておりません。
改修開始年月日
平成29年6月2日 (情報提供の開始)
効能・効果又は用途等
超音波を用いて体内の形状、性状又は動態を可視化し、画像情報を診断のために提供する。
その他
納入医療機関は全て特定しており、文書にてのご通知の上、改修を行います。