    平成22年 3月 5日作成
    

    医療機器改修の概要

    (クラスII)

  1. 一般名及び販売名

  2.   一般的名称:全身用X線CT診断装置
      販売名  :東芝スキャナ  Aquilion TSX-101A
    
    
  3. 対象ロット、数量及び出荷時期

  4.   型式名    : TSX-101A
      Ser No.   :
         2622010  ,3512057  ,3522082  ,3532101  ,3572194  ,4512060  ,4522069  ,4612009
         6BA0342138,DAA04X2003,EAA0552042,EAA0562062,EAA0572093,EAA0592138,EAA0592145,EIA05Y2214
         7BC05Z2182,HAA0612044,IAA0622009,HAA0632096,HAA0632111,HAA0632115,HAA0632129,HAA0642147
         HAA0642156,HAA0642162,HAA0652196,HAA0652204,HAA0662214,HAA0672245,HAA0692327,HAA06Y2388
         HAA06Z2471,HAA0712526,HAA0712532,HAA0722549,HAA0722550,HAA0722565,HAA0722593,HBA0732612
         HAA0732641,HAA0732642,HAA0742686,HAA0752711,HAA0782881,HAA07X2912,HAA07X2956,HAA07X2957
         HAA07X2958,HAA07Y2960,HBA07Y3004,HAA07Z3017,HAA07Z3026,HAA07Z3043,HAA07Z3055,HAA0813059
         HAA0813105,HAA0823110,HAA0823114,HAA0823124,HAA0823141,HAA0823146,HAA0833159,HAA0833160
         HAA0833162,HBA0833168,HAA0833175,HAA0833185,HBA0833189,HAA0833192,HAA0833193,HAA0833194
         HAA0833195,HAA0843199,HAA0843201,HAA0843202,HAA0843213,HAA0843215,HBA0843220,HAA0853225
         HAA0853227,HAA0853228,HBA0853229,HAA0853242,HAA0853247,HBA0863251,HBA0863258,HAA0863260
         HAA0873270,HAA0873271,HAA0873277,HAA0873278,HAA0873279,HBA0873281,HBA0873282,HAA0873283
         HAA0873295,HBA0883309,HAA0883311,HAA0883315,HAA0883316,HAA0883317,HAA0883319,HAA0883327
         HAA0883331,HAA0883335,HAA0893337,HAA0893339,HBA0893340,HAA0893346,HAA0893347,HAA0893350
         HAA0893358,HAA0893361,HAA08X3363,HAA08X3365,HAA08X3368,HAA08X3375,HAA08X3378,HAA08X3379
         HAA08X3383,HAA08Y3398,HAA08Y3413,HAA08Y3414,HBA08Y3415,HAA08Y3426,HAA08Y3429,HAA08Z3438
         HAA08Z3443,HAA08Z3450,HAA08Z3455,HAA08Z3462,HAA08Z3465,HAA08Z3468,HAB0913474,HBB0913475
         HAB0913479,HAB0913480,HAB0913483,HAB0913489,HAB0913490,HAB0913499,HAB0913502,HAB0913505
         HAB0913506,HAB0913509,HAB0913510,HBB0913512,HAB0913514,HAB0913515,HAB0913518,HAB0913521
         HAB0923528,HAB0923531,HAB0923535,HAB0923540,HAB0923541,HAB0923542,HAB0923543,HAB0923545
         HAB0923546,HAB0923547,HAB0923548,HAB0923551,HAB0923553,HAB0933556,HAB0933557,HAB0933558
         HAB0933559,HAB0933561,HBB0933565,HAB0933566,HAB0933567,HAB0933568,HAB0933569,HAB0933571
         NAA0942001,NAA0942002,NAA0942003,NBA0942004,NAA0942005,NBA0942006,NAA0942007,NAA0942008
         HAB0943590,HAB0943592,NAA0952011,NAA0952012,NBA0952014,NAA0952016,NAA0952017,NAA0952021
         NAA0962025,NAA0962027,NAA0962028,NAA0972029,NAA0972030,NAA0972035,NAA0972036,NAA0972038
         NAA0982041,NAA0982042,NAA0982045,NAA0982046,NAA0982049,NAA0982050,NAA0982051,NAA0992053
         NBA0992056,NAA0992058,NBA0992059,NAA0992061,NBA0992062,NBA0992063,NAA0992064,NAA0992065
         NAA0992066,NAA0992067,NAA0992069,NAA0992070,NBA0992071,NBA0992073,NAA09X2078,NAA09Y2085
         NAA09Y2087,NAA09Y2088,NAA09Y2090,NAA09Y2093,NAA09Y2095,NAA09Y2096,NAA09Y2097,NAA09Y2098
         NAA09Y2099,NAA09Z2100,NBA09Z2101,NAA09Z2102,NAA09Z2105,NBA09Z2107,NAA09Z2108,NBA09Z2109
         NBA1012113,NAA1022114,NAA1022116,NBA1022117
    
      数量     :244台
      出荷時期   :平成14年11月〜平成22年2月
    
    
  5. 製造販売業者等名称

  6.   製造販売業者の名称  :東芝メディカルシステムズ株式会社
      製造販売業者の住所  :栃木県大田原市下石上1385番地
      薬事の業態      :第一種医療機器製造販売業
      製造販売業許可番号  :09B1X00003
      製造所の名称     :東芝メディカルシステムズ株式会社
      製造所の住所     :栃木県大田原市下石上1385番地
      薬事の業態      :医療機器製造業
      製造業許可番号    :09BZ006008
    
    
  7. 改修理由

  8.   当該システムのオプションソフトウェアであるバリアブルピッチヘリカルスキャンシステム(寝台移動速度の
      異なる2種類のヘリカルスキャンを1回のヘリカルスキャンで撮影することができる機能)において、ソフト
      ウェアの問題により設定した寝台移動速度で撮影が行われず、結果として意図した撮影ができない可能性が
      あることが判明いたしました。
      当社としては、上記問題の対策を改修として実施することにいたしました。
    
    
  9. 危惧される具体的な健康被害

  10.   本事象により、意図した撮影ができず再検査が必要となる可能性があります。その場合、不要被曝が想定さ
      れますが、患者が受ける被曝量は通常の一回のスキャン相当であり、健康被害発生の可能性は低いと考えて
      おります。
      なお、今までに上記の問題による健康被害発生の報告は受けておりません。
    
    
  11. 改修開始年月日

  12.   平成22年 3月 5日
    
    
  13. 効能・効果又は用途等

  14.   CT装置はX線管と検出器を対向に配置して、被検者の周りを回転走査させることによって、頭部を含めた全身
      の横断断層像を得る装置で、医師もしくは診療放射線技師等の有資格者が使用します。
    
    
  15. その他

  16.   納入しました医療機関は全て特定しており、全対象装置に対策を改修として実施します。
    
    