    平成23年10月28日作成
    

    医療機器改修の概要

    (クラスII)

  1. 一般名及び販売名

  2.   一般的名称: 全身用X線CT診断装置
      販売名  : 東芝スキャナ Aquilion TSX-101A
    
    
  3. 対象ロット、数量及び出荷時期

  4.   型式名 :TSX-101A
      Ser.No :A2522003 ,A2622010 ,A3512057 ,A3572194 ,A3602021 ,B4512060 ,B4522069 ,A4542014
           A4542018 ,B4582089 ,A4622024 ,B4622204 ,B5522255 ,EAA0552042,EAA0572093,DAA0572133
           EAA0592138,IAA0622009,HAA0632115,HAA0642147,HAA0652178,HAA0652196,HAA0672245,HAA0692316
           HAA0692333,HAA0692336,HAA0692340,HAA0712526,HAA0712527,HBA0732628,HAA0732641,HAA0752711
           HAA0782870,HAA0792889,HAA0813059,HAA0813100,HAA0823110,HAA0823146,HAA0833167,HAA0843215
           HBA0853233,HAA0873277,HBA0893340,HAA0893347,HAA0893350,HAB0913480,HAB0913483,HAB0913505
           HAB0913509,HBB0913512,HAB0913514,HAB0913515,HAB0923531,HAB0923535,HAB0923540,HAB0923541
           HAB0923542,HAB0923545,HAB0923546,HAB0923547,HAB0923548,HAB0923551,HAB0923553,HAB0933556
           HAB0933557,HAB0933558,HAB0933559,HAB0933561,HBB0933565,HAB0933566,HAB0933567,HAB0933568
           HAB0933569,HAB0933571,NAA0942001,NAA0942002,NAA0942003,NBA0942004,NAA0942005,NBA0942006
           NAA0942007,NAA0942008,HAB0943590,HAB0943592,NAA0952011,NAA0952012,NBA0952014,NAA0952016
           NAA0952017,NAA0952021,NAA0962025,NAA0962027,NAA0962028,NAA0972029,NAA0972030,NAA0972035
           NAA0972036,NAA0972038,NAA0982041,NAA0982042,NAA0982045,NAA0982046,NAA0982049,NAA0982050
           NAA0982051,NAA0992053,NBA0992056,NAA0992058,NBA0992059,NAA0992061,NAA0992061,NAA0992062
           NBA0992063,NAA0992064,NAA0992065,NAA0992066,NAA0992067,NAA0992069,NAA0992070,NBA0992071
           NBA0992073,NBA1012113,NAA1022114,NAA1022116,NBA1022117,NAA1022120,NAA1022122,NBA1022124
           NAA1022125,NAA1022126,NAA1022127,NAA1022128,NAA1022130,NAA1022131,NBA1022133,NAA1032134
           NAA1032135,NAA1032136,NAA1032137,NAA1032138,NAA1032139,NAA1032140,NAA1032142,NAA1032143
           NAA1032144,NAA1032145,NAA1032147,NAA1032148,NAA1032151,NBA1032152,NAA1032153,NAA1032154
           NAA1032155,NAA1042156,NAA1042157,NAA1042158,NAA1042159,NAA1042161,NAA1042162,NAA1042163
           NAA1042166,NAA1042169,NBA1042170,NAA1042171,NBA1042173,NBA1042174,NBA1042176,NAA1062182
           NAA1062183,NAA1062184,NAA1072193,NAA1072195,NAA1082200,NAA1082201,NBA1082202,NAA1082204
           NAA1082207,NAA1082208,NAA1082209,NAA1082210,NAA1082211,NAA1082213,NAA1092215,NAA1092216
           NAA1092217,NAA1092219,NAA1092220,NAA1092221,NAA1092222,NBA1092224,NAA1092227,NBA1092228
           NAA1092233,NAA1092235,NBA1092238,NBA1112285,NAA1112287,NAA1112289,NAA1112290,NAA1112291
           NAA1112292,NAA1122293,NAA1122294,NBA1122299,NBA1122300,NBA1122303,NAA1122304,NAA1122305
           NAA1122307,NAA1122311,NAA1132313,NAA1132314,NAA1132315,NAA1132316,NBA1132317,NAA1132321
           NAA1132323,NAA1132327,NAA1142336,NAA1142337,NAA1152342,NAA1152344,EIA05Y2214,7BC05Z2182
           HAA06Z2461,HAA07X2957,HAA07X2958,HAA07Y2960,HAA08X3383,HAA08Y3413,HAA08Z3443,HAA08Z3455
           HAA08Z3462,NAA09X2075,NAA09X2078,NAA09Y2085,NAA09Y2087,NAA09Y2088,NAA09Y2090,NAA09Y2093
           NAA09Y2095,NAA09Y2096,NAA09Y2097,NAA09Y2098,NAA09Y2099,NAA09Z2100,NBA09Z2101,NAA09Z2102
           NAA09Z2105,NBA09Z2107,NBA09Z2109,NBA10X2240,NAA10X2242,NAA10X2243,NAA10X2244,NAA10X2248
           NAA10X2249,NAA10X2250,NAA10X2251,NBA10X2252,NAA10Y2264,NAA10Y2265,NAA10Z2272,NAA10Z2273
           NAA10Z2274,NAA10Z2277,NAA10Z2278,NBA10Z2279
    
    数量     :268台
    出荷時期   :平成 14年 2月〜平成 23年 5月
    
    
  5. 製造販売業者等名称

  6.   製造販売業者の名称 : 東芝メディカルシステムズ株式会社
      製造販売業者の所在地: 栃木県大田原市下石上1385番地
      許可の種類     : 第一種医療機器製造販売業
      許可番号      : 09B1X00003
      製造所の名称    : 東芝メディカルシステムズ株式会社
      製造所の住所    : 栃木県大田原市下石上1385番地
      薬事の業態     : 医療機器製造業
      製造業許可番号   : 09BZ006008
    
    
  7. 改修理由

  8.   当該システムのオプションソフトウェアであるバリアブルピッチヘリカルスキャンシステム(寝台移動速度
      の異なる2種類のヘリカルスキャンを1回のヘリカルスキャンで撮影することができる機能)において、ス
      キャン計画時に画面に表示される予測DLP値、及びスキャン後にサマリーに記録されるDLP値が、誤った値と
      なっていることが判明しました。
      なお、DLP値と同一画面上に表示されるCTDIvol値は正しい値が表示されております。
      弊社としては、上記の問題を対策したソフトウェアのインストールを改修として実施することといたしまし
      た。
    
    
  9. 危惧される具体的な健康被害

  10.   誤ったDLP値のもとで撮影条件を決め、それを実行した場合、患者が不要な被ばくを受ける可能性はあります
      が 、正しい値との差は約25%であり危害程度は軽微であると考えます。従って、重篤な健康被害発生の可能
      性は無いと考えております。
      なお、これまでに本問題による健康被害発生の報告は受けておりません。
    
    
  11. 改修開始年月日

  12.   平成23年10月28日 (情報提供の開始)
    
    
    
  13. 効能・効果又は用途等

  14.   CT装置はX線管と検出器を対向に配置して、被検者の周りを回転走査させることによって、頭部を含めた全身
      の横断断層像を得る装置で、医師もしくは診療放射線技師等の有資格者が使用します。
    
    
  15. その他

  16.   納入しました医療機関は全て特定しており、全対象装置への対応を改修として実施いたします。
    
    