閉じる

    平成21年6月25日作成
    

    医療機器回収(改修)の概要

    (クラスII)

  1. 一般名及び販売名

  2.  一般的名称  :X線透視診断装置用電動式患者台
      販売名   :
            (1)胃集検用透視撮影台 TU-MA5N
         (2)胃集検用透視撮影台 TU-MA5plus
    
    
  3. 対象製造番号、数量及び出荷時期

  4.   (1)TU-MA5N
       ・対象製造番号:
       C130489902  C185764011  KC10020201  KC10020203  KC10021201
       KC10021202  KC10021203  KC10022201  KC10022202  KC10023201
       KC10023202  KC10023203  KC10220202  KC10377201  KC10377202
       KC10377203  KC10378201  KC10378202  KC10378203  KC10643903
       KC10726201  KC10726203  KC10932201  KC10932202  KC10932203
       KC11048201  KC11048202  KC11048203  KC11237202  KC11237203
       KC11452201  KC11452203  KC11671202  KC11671203  KC11938202
       KC11938203  KC12078901  KC12127201  KC12128202  KC12128202
       KC12129202  KC12129203  KC12379201  KC12379203  KC12567302
       KC12567303  KC12824302  KC12824303  KC12825302  KC12825303
       KC13041301  KC13041302  KC13042301  KC13042302  KC13047901
       KC13047903  KC13048903  KC13116904  KC13116906  KC13282901
       KC13282902  KC13540302  KC13540303  KC13688901  KC13688902
       KC13688903  KC13688906  KC13771301  KC13771302  KC13771303
       KC13893901  KC13893902  KC13893903  KC13893906  KC13893907
       KC13893908  KC13893909  KC14030301  KC14030302  KC14030303
       KC14030303  KC14190901  KC14190905  KC14190906  KC14259002
       KC14259901  KC14425301  KC14425302  KC14425303  KC14426301
       KC14426303  KC14426312  KC14508001  KC14508003  KC14584301
       KC14584302  KC14584303  KC14585301  KC14585303  KC14642002
       KC14642004  KC14716301  KC14716302  KC14716303  KC15063301
       KC15063303  KC15102001  KC15102002  KC15195001  KC15195002
       KC15195003  KC15195004  KC15195403  KC15388302  KC15389401
       KC15390002  KC15506404  KC15540002  KC15554001  KC15762002
       KC15763002  KC15802401  KC15802402  KC15970001  KC15970003
       KC15993401  KC15993402  KC15993403  KC15994402  KC15994403
       KC16198001  KC16235502  KC16363001  KC16363003  KC16364001
       KC16364002  KC16364003  KC16539002  KC16539003  KC16738002
       KC16738003  KC16739002  KC16739003  KC16941401  KC16941402
       KC16941403  KC16998001  KC16998103  KC17222401  KC17222402
       KC17222402  KC17222403  KC17489401  KC17489402  KC17490401
       KC17490403  KC17544101  KC17545101  KC17608101  KC17608103
       KC17616101  KC17616103  KC17757101  KC17757103  KC17796401
       KC17796402  KC17797402  KC17797403  KC17996403  KC18021103
       KC18288101  KC18288102  KC18288103  KC18440101  KC18440102
       KC18441003  KC18441101  KC18441102  KC18441103  KC18576402
       KC18577401  KC18577402  KC18577403  KC18578401  KC18578402
       KC18578403  KC18633101  KC18633102  KC18775101  KC18775102
       KC19033101  KC19033102  KC19033103  KC19211102  KC19211102
       KC19211103  KC19211103  KC19441102  KC19541101  KC19541103
       KC19542102  KC19542103  KC19910101  KC19910102  KC19910203
       KC19911201  KC19911203  KX15399406  SX11849002  SX12183002
       SX12464103  SX13220205  SX13719205  SX14754401  SX14754402
       SX14754403  SX14894403  SX15195401  SX15195401  SX15195402
       SX15399402  SX15399403  SX15399405  SX15506401  SX15506403
       SX15506405  SX15619501  SX15619503  SX15619504  SX15644501
       SX15644502  SX15644503  SX15644504  SX15644505  SX15704502
       SX15786501  SX15786502  SX15786503  SX15786505  SX15807501
       SX15807502  SX15807503  SX16008501  SX16008502  SX16008503
       SX16008504  SX16008505  SX16126502  SX16126503  SX16179501
       SX16179502  SX16179503  SX16207504  SX16235501  SX16286501
       SX16286502  SX16286503  SX16287501  SX16287503  SX16344501
       SX16344502  SX16344503  SX16344504  SX16344505  SX16397501
       SX16397502  SX16397503  SX16398601  SX16398602  SX16398604
       SX16398607  SX16453601  SX16476601  SX16476602  SX16476604
       SX16476604  SX16476605  SX16476607  SX16604603  SX16718602
       SX16718603  SX16718604  SX16718605  SX16805601  SX16805602
       SX16805603  SX16805604  SX16805605  SX16888601  SX16888602
       SX16888603  SX16888604  SX16888605  SX16957601  SX16957602
       SX16957603  SX16957604  SX16957605  SX17004603  SX17020601
       SX17020602  SX17020603  SX17020604  SX17079601  SX17079602
       SX17079603  SX17079604  SX17079605  SX17101701  SX17101702
       SX17101703  SX17101704  SX17101705  SX17145701  SX17145702
       SX17145703  SX17145704  SX17145705  SX17369701  SX17369702
       SX17369703  SX17450701  SX17450702  SX17450703  SX17450704
       SX17474702  SX17479701  SX17479703  SX17512701  SX17512702
       SX17528701  SX17528702  SX17528703  SX17545102  SX17600701
       SX17600702  SX17600703  SX17658701  SX17658702  SX17658703
       SX17712702  SX17712702  SX17712703  SX17742801  SX17742803
       SX17787801  SX17820801  SX17820802  SX17853801  SX17853802
       SX17911801  SX17911802  SX17911803  SX17911804  SX17911805
       SX17967802  SX18049801  SX18049802  SX18049803  SX18049804
       SX18049805  SX18103801  SX18103802  SX18103803  SX18163801
       SX18163802  SX18163803  SX18251801  SX18251802  SX18251803
       SX18308801  SX18308802  SX18308803  SX18361801  SX18361802
       SX18361803  SX18422801  SX18422802  SX18422803  SX18479802
       SX18479804  SX18479805  SX18479813  SX18549901  SX18549902
       SX18549903  SX18607902  SX18607903  SX18774901  SX18774902
       SX18774903  SX18774904  SX18774905  SX18779901  SX18779902
       SX18779903  SX18874901  SX18874902  SX18874903  SX50361502
       SX51521767  SX52452802  SX53780802  SX54619902  SX56846524
       SX56847524  SX57214503
       ・出荷数量:407台
       ・出荷時期:平成1年3月31日 〜 平成11年12月21日
      (2)TU-MA5plus
    ・対象製造番号:SX18964901  SX51547401  SX57363101  SX57452202
    ・出荷数量:4台
    ・出荷年月日  :平成11年12月21日  〜  平成16年3月31日
    
    
  5. 製造販売業者等名称

  6.   製造販売業者の名称  :株式会社日立メディコ
      製造販売業者の住所  :東京都千代田区外神田四丁目14番1号
      主たる機能を有する  :千葉県柏市新十余二2番地1
      事業所所在地
      薬事の業態        :第一種医療機器製造販売業者
      製造販売業許可番号  :12B1X00004
      製造所の名称      :
       (1) TU-MA5N     :株式会社 日立メディコ 柏事業場
                   :株式会社 日立メディコ 大阪工場
       (2) TU-MA5plus    :株式会社 日立メディコ 大阪工場
      製造所の住所      :
       (1) TU-MA5N     :千葉県柏市新十余二2番地1
                     大阪府西成区長橋二丁目6番37号
       (2) TU-MA5plus    :大阪府西成区長橋二丁目6番37号
      薬事の業態        :
       (1) TU-MA5N     :医療機器製造業
       (2) TU-MA5plus    :医療機器製造業
      製造業許可番号    :
       (1) TU-MA5N     :12BZ000058(柏事業場)
                     27BZ000662(大阪工場)
       (2) TU-MA5plus    :27BZ000662
    
    
  7. 回収(改修)理由

  8.  国内の納入先において、臥位にて検査を受けていた被検者の足先が被検者を載せるための天板の端部にあった
    隙間に入りこみ、その後の体位変換の際、指先に過大な力が加わったために右足薬指の骨にひびが入った事例が
    発生したとの報告を受けました。
     調査の結果、構造上天板端部には幅1cm深さ1.5cmほどの隙間があり、通常の検査では、この隙間に足先が入
    り、健康被害が発生するほどの力が掛かるような状態となる可能性は低いものと考えておりましたが、今回健康
    被害の発生が確認されたため、当該の隙間をふさぐ部品を取り付けることで健康被害の発生を防止する回収(改
    修)を実施することと致しました。
    
    
  9. 危惧される具体的な健康被害

  10.  検査中の体位変換の途中で足先が当該の隙間に入り込んだ場合には、指に大きな力が加わり、骨にひびが入る
    等の健康被害が発生する可能性があります。
     なお、機器使用中の被検者の様子は常時医師や診療放射線技師等が監視しているため、今回の事象により重篤
    な健康被害が発生することはまず考えられません。
    
    
  11. 回収(改修)開始年月日

  12.    平成21年6月20日
    
    
  13. 効果・効能又は用途等

  14.  本装置は、X線を発生させて撮影を行い、フィルム等に人体透過後の画像を投影させ、検査、診断に利用する
    装置で、医師又は診療放射線技師等の有資格者が使用します。
    
    
  15. その他

  16.    対象医療機関の名称と所在地は全て把握しております。
    
    
  17. 担当者及び連絡先

  18.   担当者名  :品質保証本部 薬事情報管理部  平出博一  池谷和之
      連絡先  :株式会社  日立メディコ 柏事業場
       千葉県柏市新十余二2番地1
       電話  04-7131-4167
          FAX  04-7131-4774