一覧へ戻る |
一般名及び販売名
一般的名称:全身用X線CT診断装置 販 売 名:東芝スキャナ Asteion TSX-021A
対象ロット、数量及び出荷時期
形式名 : TSX-021A SerNo. : 0512006,0512007,0512008,0512009,0512010,0512061,0512090,0512092 0512093,0512094,0522011,0522012,0522013,0522014,0522015,0522016 0522017,0522018,0522019,0522084,0522085,0522098,0522099,0522100 0522102,0532002,0532021,0532025,0532027,0532028,0532029,0532030 0532031,0532069,0532087,0532089,0532103,0532104,0532108,0532109 0532110,0532111,0532112,0532113,0532114,0542036,0542037,0542116 0552041,0552090,0562035,0562046,0562048,0562049,0562050,0562051 0562121,0572053,0582001,0582002,0582057,0582059,0582060,0582063 0582064,0582065,0582069,0582126,0582129,0582130,0582131,0582133 0592003,0592004,0592005,0592006,0592007,0592008,0592009,0592043 0592135,0592136,0592137,0592138,0602010,0602072,0602074,0602076 0602077,0602078,0612011,0612012,0612013,0612083,0612085,0612100 0612143,0612144,0622014,0622015,0622080,0622086,0622106,0622145 0622146,0622147,0622148,1512016,1512017,1512018,1512019,1512020 1512021,1512022,1512023,1512094,1512097,1512149,1512152,1522024 1522025,1522026,1522027,1522029,1522030,1522031,1522032,1522098 1522101,1522102,1522103,1522104,1522106,1522153,1522154,1522156 1522157,1522158,1522159,1532001,1532034,1532035,1532037,1532038 1532039,1532040,1532041,1532042,1532043,1532044,1532079,1532108 1532110,1532111,1532113,1532160,1532161,1532163,1532169,1532170 1532171,1532172,1532173,1532174,1542045,1542046,1552048,1552049 1552122,1552123,1562050,1562051,1562052,1562127,1572002,1572053 1572054,1572055,1572056,1572057,1572137,1582058,1582059,1582060 1582061,1582062,1582142,1582145,1592006,1592007,1592047,1592063 1592064,1592065,1592066,1592067,1592068,1592069,1592070,1592071 1592149,1592150,1592151,1602072,1602156,1602158,1612073,1612074 1612075,1612076,1612077,1612166,1612188,1622078,1622079,1622082 1622083,1622084,1622085,1622127,2512015,2512176,2512190,2522018 2522155,2522178,2522193,2522195,2532019,2532020,2532021,2532185 2532188,2532196,2542197,2542198,2542199,2542200,2552201,2562024 2562134,2572025,2572226,2582231,2582237,2592029,2592199,2592243 2592244,2592245,2602249,2612203,2612246,2612257,2622209,2622210 2622258,2622268,3512033,3512272,3522212,3522214,3522227,3522285 3522288,3522290,3532215,3532216,3532217,3542141,3542310,3572329 3572331,3592342,3602041,3612219,3622144,3622220,3622365,3JB0572402 3JB0612404,4512145,4512376,4522043,4552147,4592221,4602222,4602399 4622012,4622223,4622224,5512400,5522401,9512001,9512003,9522004 9522005,9522006,9522007,9522010,9522011,9522013,9522014,9522015 9522016,9522017,9522021,9532021,9532022,9532023,9532024,9532025 9532026,9532027,9562037,9572042,9572044,9572046,9572049,9572050 9572051,9582051,9582052,9582053,9582054,9582055,9582057,9592059 9592062,9592063,9592065,9592066,9592067,9592068,9602066,9602073 9602074,9612072,9612073,9612075,9612076,9612077,9622003,9622004 9622082,9622087 数量 :346台 出荷時期:平成11年 1月〜平成18年 1月
製造販売業者等名称
製造販売業者の名称 :東芝メディカルシステムズ株式会社 製造販売業者の住所 :栃木県大田原市下石上1385番地 薬事の業態 :第一種医療機器製造販売業 製造販売業許可番号 :09B1X00003 製造所の名称 :東芝メディカルシステムズ株式会社 製造所の住所 :栃木県大田原市下石上1385番地 薬事の業態 :医療機器製造業 製造業許可番号 :09BZ006008
改修理由
架台に搭載されているX線高圧発生ユニットにおいて放熱処理が不十分であったため、故障する可能 性があることが判りました。 X線高圧発生ユニット内に使用している素子はX線曝射時に発熱します。発生した熱は放熱板に逃がす 構造になっておりますが、放熱部材の特性に問題があり、その結果、素子の温度が上昇し、長期間の 使用により破損する可能性があります。また、素子の破損状況によっては、破損した後でも素子に電 源が供給される場合があり、素子の周囲の部品に損傷がおよぶ可能性があります。 当社としては、十分な放熱ができる様に対策したユニットに交換し、さらに素子が破損した場合でも 確実に電源を遮断できるように、架台電源入力部のブレーカを漏電検出機能付きブレーカに交換する 対策を改修として実施することといたしました。 素子の破損時には、異音の発生と発煙の可能性がありますが、上記の対策を行うことにより放熱処理 に起因した素子破損を抑制できるものと考えております。
危惧される具体的な健康被害
素子が破損しても、不要被曝が発生することはありません。 素子の破損状態によっては、素子の周囲の部品に損傷がおよぶ可能性がありますが、架台内の他のユ ニットや架台外への影響は無いと考えております。 上記より、患者様への身体的な被害の可能性は無いと考えております。
改修開始年月日
平成19年 9月 4日
効能・効果又は用途等
CT装置はX線管と検出器を対向に配置して、被検者の周りを回転走査させることによって、頭部を含 めた全身の横断断層像を得る装置で、医師もしくは診療放射線技師が使用します。
その他
納入しました医療機関は全て特定しており、全対象装置に対策を改修として実施します。
担当者名及び連絡先
担当者氏名 :薮内 良昌、中原 朋哉 連絡先 :東芝メディカルシステムズ株式会社 栃木県大田原市下石上1385番地 Tel:0287-26-6211 Fax:0287-26-6050