一覧へ戻る

平成19年 9月 4日作成

医療機器改修の概要

(クラスII)



  1. 一般名及び販売名

      一般的名称:全身用X線CT診断装置
      販 売 名:東芝スキャナ Asteion TSX-021A
    
    
  2. 対象ロット、数量及び出荷時期

      形式名 : TSX-021A
      SerNo. : 0512006,0512007,0512008,0512009,0512010,0512061,0512090,0512092
           0512093,0512094,0522011,0522012,0522013,0522014,0522015,0522016
           0522017,0522018,0522019,0522084,0522085,0522098,0522099,0522100
           0522102,0532002,0532021,0532025,0532027,0532028,0532029,0532030
           0532031,0532069,0532087,0532089,0532103,0532104,0532108,0532109
           0532110,0532111,0532112,0532113,0532114,0542036,0542037,0542116
           0552041,0552090,0562035,0562046,0562048,0562049,0562050,0562051
           0562121,0572053,0582001,0582002,0582057,0582059,0582060,0582063
           0582064,0582065,0582069,0582126,0582129,0582130,0582131,0582133
           0592003,0592004,0592005,0592006,0592007,0592008,0592009,0592043
           0592135,0592136,0592137,0592138,0602010,0602072,0602074,0602076
           0602077,0602078,0612011,0612012,0612013,0612083,0612085,0612100
           0612143,0612144,0622014,0622015,0622080,0622086,0622106,0622145
           0622146,0622147,0622148,1512016,1512017,1512018,1512019,1512020
           1512021,1512022,1512023,1512094,1512097,1512149,1512152,1522024
           1522025,1522026,1522027,1522029,1522030,1522031,1522032,1522098
           1522101,1522102,1522103,1522104,1522106,1522153,1522154,1522156
           1522157,1522158,1522159,1532001,1532034,1532035,1532037,1532038
           1532039,1532040,1532041,1532042,1532043,1532044,1532079,1532108
           1532110,1532111,1532113,1532160,1532161,1532163,1532169,1532170
           1532171,1532172,1532173,1532174,1542045,1542046,1552048,1552049
           1552122,1552123,1562050,1562051,1562052,1562127,1572002,1572053
           1572054,1572055,1572056,1572057,1572137,1582058,1582059,1582060
           1582061,1582062,1582142,1582145,1592006,1592007,1592047,1592063
           1592064,1592065,1592066,1592067,1592068,1592069,1592070,1592071
           1592149,1592150,1592151,1602072,1602156,1602158,1612073,1612074
           1612075,1612076,1612077,1612166,1612188,1622078,1622079,1622082
           1622083,1622084,1622085,1622127,2512015,2512176,2512190,2522018
           2522155,2522178,2522193,2522195,2532019,2532020,2532021,2532185
           2532188,2532196,2542197,2542198,2542199,2542200,2552201,2562024
           2562134,2572025,2572226,2582231,2582237,2592029,2592199,2592243
           2592244,2592245,2602249,2612203,2612246,2612257,2622209,2622210
           2622258,2622268,3512033,3512272,3522212,3522214,3522227,3522285
           3522288,3522290,3532215,3532216,3532217,3542141,3542310,3572329
           3572331,3592342,3602041,3612219,3622144,3622220,3622365,3JB0572402
           3JB0612404,4512145,4512376,4522043,4552147,4592221,4602222,4602399
           4622012,4622223,4622224,5512400,5522401,9512001,9512003,9522004
           9522005,9522006,9522007,9522010,9522011,9522013,9522014,9522015
           9522016,9522017,9522021,9532021,9532022,9532023,9532024,9532025
           9532026,9532027,9562037,9572042,9572044,9572046,9572049,9572050
           9572051,9582051,9582052,9582053,9582054,9582055,9582057,9592059
           9592062,9592063,9592065,9592066,9592067,9592068,9602066,9602073
           9602074,9612072,9612073,9612075,9612076,9612077,9622003,9622004
           9622082,9622087
    
       数量  :346台
      出荷時期:平成11年 1月〜平成18年 1月
    
    
  3. 製造販売業者等名称

      製造販売業者の名称 :東芝メディカルシステムズ株式会社
      製造販売業者の住所 :栃木県大田原市下石上1385番地
      薬事の業態     :第一種医療機器製造販売業
      製造販売業許可番号 :09B1X00003
      製造所の名称    :東芝メディカルシステムズ株式会社
      製造所の住所    :栃木県大田原市下石上1385番地
      薬事の業態     :医療機器製造業
      製造業許可番号   :09BZ006008
    
    
  4. 改修理由

      架台に搭載されているX線高圧発生ユニットにおいて放熱処理が不十分であったため、故障する可能
      性があることが判りました。
      X線高圧発生ユニット内に使用している素子はX線曝射時に発熱します。発生した熱は放熱板に逃がす
      構造になっておりますが、放熱部材の特性に問題があり、その結果、素子の温度が上昇し、長期間の
      使用により破損する可能性があります。また、素子の破損状況によっては、破損した後でも素子に電
      源が供給される場合があり、素子の周囲の部品に損傷がおよぶ可能性があります。
      当社としては、十分な放熱ができる様に対策したユニットに交換し、さらに素子が破損した場合でも
      確実に電源を遮断できるように、架台電源入力部のブレーカを漏電検出機能付きブレーカに交換する
      対策を改修として実施することといたしました。
      素子の破損時には、異音の発生と発煙の可能性がありますが、上記の対策を行うことにより放熱処理
      に起因した素子破損を抑制できるものと考えております。
    
    
  5. 危惧される具体的な健康被害

      素子が破損しても、不要被曝が発生することはありません。
      素子の破損状態によっては、素子の周囲の部品に損傷がおよぶ可能性がありますが、架台内の他のユ
      ニットや架台外への影響は無いと考えております。
      上記より、患者様への身体的な被害の可能性は無いと考えております。
    
    
  6. 改修開始年月日

      平成19年 9月 4日
    
    
  7. 効能・効果又は用途等

      CT装置はX線管と検出器を対向に配置して、被検者の周りを回転走査させることによって、頭部を含
      めた全身の横断断層像を得る装置で、医師もしくは診療放射線技師が使用します。
    
    
  8. その他

      納入しました医療機関は全て特定しており、全対象装置に対策を改修として実施します。
    
    