平成19年2月6日作成

医療機器回収(改修)の概要

(クラスII)



  1. 一般名及び販売名

    
    一般的名称          :X線透視診断装置用電動式患者台
    
    販売名             :カセッテレスX線透視撮影装置 TU-130
    
    
    
    
  2. 対象製造番号、数量及び出荷時期

    
    対象製造番号
    
    KC10065601    KC10065602    KC10066601    KC10066602    KC10067601
    
    KC10067602    KC10090301    KC10090302    KC10091301    KC10091302
    
    KC10300601    KC10300602    KC10301601    KC10301602    KC10313301
    
    KC10313302    KC10450302    KC10479601    KC10479602    KC10480601
    
    KC10480602    KC10661302    KC10718601    KC10718602    KC10719601
    
    KC10719602    KC10720601    KC10720602    KC10922301    KC10922302
    
    KC10923301    KC10923302    KC10924301    KC10924302    KC10925301
    
    KC10925302    KC10926301    KC10926302    KC11045601    KC11045602
    
    KC11046601    KC11046602    KC11455301    KC11455302    KC11494301
    
    KC11494302    KC11495301    KC11495302    KC11620301    KC11625301
    
    KC11625302    KC11739001    KC11739002    KC11748101    KC11748102
    
    KC11864402    KC12157301    KC12157302    KC12158401    KC12158402
    
    KC12271401    KC12271402    KC12272401    KC12272402    KC12316401
    
    KC12316402    KC12317401    KC12317402    KC12447401    KC12447402
    
    KC12448401    KC12448402    KC12449401    KC12449402    KC12450401
    
    KC12450402    KC12451401    KC12451402    KC12452401    KC12452402
    
    KC12453401    KC12453402    KC12454401    KC12454402    KC12455401
    
    KC12455402    KC12456401    KC12456402    KC12572101    KC12572102
    
    KC12572103    KC12572104    KC12573101    KC12573102    KC12573103
    
    KC12573104    KC12574101    KC12574102    KC12574103    KC12574104
    
    KC12575101    KC12575102    KC12575103    KC12575104    KC12696401
    
    KC12696402    KC12697401    KC12697402    KC12698401    KC12698402
    
    KC12763401    KC12763402    KC12764401    KC12764402    KC12865401
    
    KC12865402    KC12866401    KC12866402    KC12950401    KC12950402
    
    KC12951401    KC12951402    KC13252402    KC13253401    KC13253402
    
    KC13254401    KC13254402    KC13271102    KC13271103    KC13271104
    
    KC13382401    KC13382402    KC13383401    KC13383402    KC13384401
    
    KC13384402    KC13547101    KC13547103    KC13583402    KC13584401
    
    KC13584402    KC13697102    KC13697104    KC14017101    KC14017102
    
    KC14017103    KC14017104    KC14107401    KC14108401    KC14108402
    
    KC14109401    KC14212401    KC14212402    KC14280101    KC14280102
    
    KC14280103    KC14641401    KC14641402    KC14642401    KC14642402
    
    KC14708401    KC14708402    KC14845501    KC14845502    KC14846501
    
    KC14846502    KC15100501    KC15100502    KC15101502    KC15102501
    
    KC15102502    KC15237201    KC15237204    KC15354501    KC15354502
    
    KC15355501    KC15355502    KC15377201    KC15377202    KC15377204
    
    KC15509501    KC15509502    KC15510501    KC15510502    KC15586201
    
    KC15586203    KC15886201    KC15886203    KC15888501    KC15888502
    
    KC15888504    KC16103201    KC16183502    KC16184501    KC16184502
    
    KC16185501    KC16185502    KC16305201    KC16305202    KC16306201
    
    KC16306202    KC16513501    KC16514501    KC16514502    KC16907501
    
    KC16907502    KC16908501    KC16908502    KC17041501    KC17530201
    
    KC17530202    KC17782501    KC17782502    KC17821501    KC17821502
    
    KC17822502    KC17931201    KC17931202    KC18246302    KC18247301
    
    KC18247302    KC18278501    KC18278502    KC18279501    KC18279502
    
    KC18366301    KC18366302    KC18529601    KC18529602    KC18530601
    
    KC18530602    KC18763501    KC18802601    KC18802602    KC18803601
    
    KC18803602    KC18804601    KC18804602    KC18805601    KC18805602
    
    KC18897301    KC18897302    KC18969301    KC18972301    KC18972302
    
    KC18973301    KC19137601    KC19137602    KC19138601    KC19138602
    
    KC19139601    KC19139602    KC19140601    KC19140602    KC19141601
    
    KC19141602    KC19142601    KC19142602    KC19288301    KC19288302
    
    KC19289302    KCR2001201
    
    
    
    出荷数量:262台
    
    出荷時期:平成13年3月20日 〜 平成18年6月18日
    
    
    
    
  3. 製造販売業者等名称

    
    製造販売業者の名称    :株式会社 日立メディコ
    
    製造販売業者の住所    :東京都千代田区外神田四丁目14番1号
    
    薬事の業態            :第一種医療機器製造販売業者
    
    製造販売業許可番号    :12B1X00004
    
    製造所の名称          :株式会社 日立メディコ 柏事業場
    
    製造所の住所          :千葉県柏市新十余二2番地1
    
    薬事の業態            :医療機器製造業
    
    製造業許可番号        :12BZ000058
    
    
    
    
  4. 回収(改修)理由

    
    国内の納入先において検査中に被検者を載せる天板の意図しない動作が発生し天板が上方
    
    向に移動したため、操作中の技師が機器を停止させたという事例が報告されました。
    
    原因は、上下方向に天板を移動させるためのモータを駆動する制御回路の一部が故障しモー
    
    タに対し誤った制御信号が出力されたために意図しない動作が発生したことが判りました。
    
    そこで、制御回路の一部を変更し制御回路が故障した場合であってもモータに対し誤った
    
    制御信号が送られないような回路に変更する回収(改修)を実施することと致しました。
    
    なお、本件による健康被害発生の報告は受けておりません。
    
    
    
    
  5. 危惧される具体的な健康被害

    
    天板を制御するモータに対し誤った制御信号が出力され天板の動作に異常が発生した場合
    
    でも、検査中被検者は握り棒を握っているため、被検者の手が天板の外側に出ることはあ
    
    りません。(握り棒を握ることは、警告銘板及び取扱説明書上に記載されています。)
    
    また、被検者の手足は天板上から外部に出ないようサイドガード等により保護されており、
    
    かつ被検者の様子は常時技師又は医師が監視、確認しているため、今回の事象により健康
    
    被害が発生することはまず考えられません。
    
    
    
    
  6. 回収(改修)開始年月日

    
    平成19年1月23日
    
    
    
    
  7. 効果・効能又は用途等

    
    本装置は、X線を発生させて撮影を行い、フィルム等に人体透過後の画像を投影させ、検
    
    査、診断に利用する装置で、医師又は診療放射線技師等の有資格者が使用します。
    
    
    
    
  8. その他

    
    対象医療機関の名称と所在地は全て把握しております。
    
    
    
    
  9. 担当者及び連絡先

    
    担当者名    :薬事統括本部    平出博一    池谷和之
    
    連絡先      :株式会社         日立メディコ 柏事業場
    
    千葉県柏市新十余二2番地1
    
    電話     04-7131-4167
    
    FAX      04-7131-4774