一覧へ戻る |
一般名及び販売名
一般的名称 :天井取付け式X線管支持器 販売名 :医療用エックス線装置及び医療用エックス線装置用エックス線管
対象製造番号、数量及び出荷時期
対象製造番号 : (1) SX−V7 KC13979601 KC17762701 (2) SX−A7 KC10013702 KC10013704 KC10013705 KC10013706 KC10013707 KC10013708 KC10013711 KC10013712 KC10013713 KC10013714 KC10013715 KC10013717 KC10013718 KC10013719 KC10013720 KC10245701 KC10245702 KC10245708 KC10245709 KC10245711 KC10245714 KC10245715 KC10403701 KC10403702 KC12530508 KC12812507 KC12852503 KC13167503 KC13167504 KC13362504 KC13362508 KC13362511 KC13362512 KC13362513 KC13362515 KC13602502 KC13602503 KC13602504 KC13602507 KC13603501 KC13603504 KC13603507 KC13603508 KC13603510 KC13603511 KC13603512 KC13933602 KC13933605 KC13934601 KC13934601 KC13934602 KC13934603 KC13934604 KC13934605 KC13934606 KC13934608 KC13934609 KC13934610 KC13934611 KC13934614 KC13934615 KC14272602 KC14272603 KC14272604 KC14272607 KC14272608 KC14272610 KC14272612 KC14272613 KC14272614 KC14272615 KC14272616 KC14512601 KC14512602 KC14512603 KC14512607 KC14512608 KC14512609 KC14512610 KC14512611 KC14512612 KC14512613 KC14512614 KC14512615 KC14512616 KC14512617 KC14530601 KC14530603 KC14530605 KC14530606 KC14530607 KC14530609 KC14530614 KC14773601 KC14773602 KC14773604 KC14773605 KC14773606 KC14773607 KC14773608 KC14773612 KC15074601 KC15074602 KC15074603 KC15074606 KC15074607 KC15074608 KC15074609 KC15074610 KC15074613 KC15074614 KC15074615 KC15212601 KC15212602 KC15212604 KC15212605 KC15212606 KC15212608 KC15212609 KC15212610 KC15416601 KC15416602 KC15416606 KC15416608 KC15416610 KC15416611 KC15416613 KC15416615 KC15647602 KC15647603 KC15647606 KC15647607 KC15647608 KC15647615 KC16020601 KC16020602 KC16257601 KC16257605 KC16257606 KC16257607 KC16257608 KC16257611 KC16257612 KC16562601 KC16562602 KC16562603 KC16562604 KC16562605 KC16562606 KC16562607 KC16562608 KC16562613 KC16562614 KC16562615 KC16855601 KC16855602 KC16855604 KC16855606 KC16855607 KC16855608 KC16855609 KC16855610 KC16855612 KC16855613 KC16855615 KC17120601 KC17120602 KC17120603 KC17120604 KC17121601 KC17121602 KC17121603 KC17121604 KC17121605 KC17121606 KC17121607 KC17121608 KC17121609 KC17121612 KC17121612 KC17121613 KC17121614 KC17121615 KC17121616 KC17121617 KC17121618 KC17122701 KC17122702 KC17493701 KC17493702 KC17493703 KC17493704 KC17493705 KC17493706 KC17493708 KC17493709 KC17493710 KC17493711 KC17493712 KC17493713 KC17493715 KC17493719 KC18041702 KC18041704 KC18041705 KC18041707 KC18041708 KC18041710 KC18042702 KC18042703 KC18042705 KC18042706 KC18042709 KC18042710 KC18042713 KC18042714 KC18042715 KC18042719 KC18042720 KC18392701 KC18392703 KC18392704 KC18392705 KC18392706 KC18392707 KC18392708 KC18392712 KC18392714 KC18392715 KC18392716 KC18392717 KC18392719 KC18392720 KC18789701 KC18789702 KC18789704 KC18789705 KC18789706 KC18789707 KC18789708 KC18789713 KC18789717 KC18789718 KC19094701 KC19094702 KC19094703 KC19094704 KC19094705 KC19094706 KC19094707 KC19094708 KC19094709 KC19094710 KC19094711 KC19094713 KC19094714 KC19094716 KC19094717 KC19094718 KC19094719 KC19094720 KC19183701 KC19183704 KC19183705 KC19183706 KC19183707 KC19183708 KC19183709 KC19183710 KC19183713 KC19183714 KC19183715 KC19183716 KC19183717 KC19183718 KC19183719 KC19183720 出荷数量 : (1) SX−V7: 2台 (2) SX−A7:277台 出荷時期 : (1)SX−V7:平成8年12月27日 〜 平成9年 3月17日 (2)SX−A7:平成7年10月23日 〜 平成9年12月19日
製造販売業者等名称
製造販売業者の名称 :株式会社 日立メディコ 製造販売業者の住所 :東京都千代田区外神田四丁目14番1号 薬事の業態 :第一種医療機器製造販売業者 製造販売業許可番号 :12B1X00004 製造所の名称 :株式会社 日立メディコ 柏事業場 製造所の住所 :千葉県柏市新十余二2番地1 薬事の業態 :医療機器製造業 製造業許可番号 :12BZ000058
回収(改修)理由
国内の納入先において機器を使用するための準備作業中、天井に取り付けられている横移 動機構部内にあるX線管支持器の位置を固定するブレーキ部分から異臭と少量の発煙が発 生したため機器の使用を中止した事例が報告されました。 本機器は、検査室の天井に取り付けられたレールからX線管支持器本体が吊り下げられて おり、機器は人力によりレール上を移動させることができます。検査時は、任意の位置で X線管支持器内に設けられたブレーキを掛けることで停止させることができます。 調査の結果、ブレーキ本体とレール間の接触状態が適切でなかったためにブレーキ機構の 放熱効果が低下し、その状態のままで機器を長期間使用したためにブレーキ内部にあるコ イルの絶縁状態が熱により劣化してレアショート(短絡)が発生して異臭・発煙に至った ことが原因と判明しました。 この現象を防止するため、レアショートが発生した場合に発生する過大な電流を検出しブ レーキ回路への電力供給を遮断する回路に変更する回収(改修)を実施することと致しま した。 なお、本件による健康被害発生の報告は受けておりません。
危惧される具体的な健康被害
ブレーキコイルのレアショートが発生するとコイル内部の加熱により異臭や発煙が発生す ることがありますが、機器に使用されている部品は難燃性もしくは自己消火性があるため 延焼、類焼に至る可能性は、まず考えられません。 また、発生する異臭、発煙により健康被害が発生する可能性は、まず考えられません。
回収(改修)開始年月日
平成18年12月5日
その他
本機器は、本装置はX線撮影装置の一部として、X線管球等を検査室天井から懸架支持し、 手動により移動させるもので、医師もしくは診療放射線技師が使用します。
その他
対象医療機関の名称と所在地は全て把握しております。
担当者及び連絡先
担当者名 :薬事統括本部 平出博一 池谷和之 連絡先 :株式会社 日立メディコ 柏事業場 千葉県柏市新十余二2番地1 電話 04-7131-4167 FAX 04-7131-4774