一般名及び販売名
一般的名称:X線用TV装置 販売名 :X線TVカメラ MTV-500A
対象ロット、数量及び出荷時期
型式 :MTV-500A SerNo. :6592223,7572571,8512774,8512739,7602675,7602676,8552854,8582913,8582920,8552865 8582916,8522797,7602671,8572907,8572904,7562540,7562541,7612684,7612687,8582923 8582912,8522789,8512740,8512743,8512747,8532815,7522389,8562875,6512003,6572146 8522794,8522808,8512744,8512746,8512767,7572575,8582921,8582922,6572170,7592646 7592648,6562123,6562124,8512755,8582924,8532814,8522798,7602665,7602678,7602680 7602666,9622117,8602016,1532286,0522152,0592212,8592003,8592004,2552434,2522386 0512137,9512030,0512136,0522141,1572312,2532405,9552067,5222497,0562190,0562189 2532406,1512257,1512254,9552065,9582088,9572078,9572079,1572309,9532052,9532053 9532049,1512262,1562306,2552430,0592215,8592012,2592461,1532289,0622239,1572311 1522279,1542300,3532487,3532486,2522387,0542172,0602226,9622117,2522381,0522153 2522401,0532154,1612356,0542173,9552069,0512135,8602014,9582087,3512481,0552181 9582083,9512023,9512024,1522278,2592462,0532156,9612113,0602222,9522034,2602465 2522392,1572326,1572327,2522385,1532288,9592096,9622121,0512122,1522270,0612230 0542177,9522043,0532162,9542063,9542064,1522268,1522269,0592210,2522400,2522399 0612235,1592331,2512372,2542413,1532287,9522046,8622017,8622018,0592219,0512134 2532407,2532408,9592102,2592459,9597003,0592218,8622019,0552182,0592221,2522384 0532159,9612109,9612108,3522483,0522148,1522273,0582207,0542175,2532403,9582084 9582085,2522398,0542169,9592094,9582090,9582089,1512261,0612237,9572076,2602487 9512029,1622365,1532284,0572193,0572194,0512132,0512133,1572328,8592006,9532050 1512264,2522394,0582201,0582204,2562439,0622244,0622245,9522047,9532056,2572448 0582208,9582082,1512253,0612233,0532158,1572324,1572323,2602466,0522143,1512251 9592008,0582209,9572077,1572316,1572317,8592007,9522040,9522041,1572308,2532409 1532285,1622369,0572197,1522239,2542416,0532157,1602342,9552070,8592013,2532410 9522038,9522039,9512026,0622241,1592330,9602105,0592214,9572080,0622246,1612361 1612362,1542297,0522150,2562443,2562444,9532051,2552433,9562071,1622368,2522397 9592104,1612860,9522033,9522036,9512028,9512027,0512138,9512031,8622020,0542174 9592091,9622114,8592011,2522389,2522388,0592213,0602227,8592009,9592098,9512032 0512127,0512128,0522142,1542293,9592092,2522369,2512380,0602223,9612110,9512025 9522045,0572195,9572073,2532404,1512252,0512126,9542062,9522035,8592008,8592002 8592001,9592097,9552066,5222721,9587001,9532005,1622363,4552496,9582086,9572081 1592334,0582205,9612107,1512265,1512266,0522140,9532054,9572075,1592332,2562422 2562442,1522111,1512260,2602468,0512129,9592103,9612112,2522382,2522396,2542414 9587002,8602015,0622249,0612232,2582455,1512256,0522147,9532061,2612384,2512314 8592021,8592022,2552348,0552175,0582124,0572189,0512144,0512145,1592286,1592287 9622131,9622132,8552006,2532330,2522324,0562182,0562183,1512228,2542333,2542332 8592025,8592024,2592376,9582099,1572272,3522396,3522395,9562083,2562355,1532250 0542158,0542159,8537008,8537007,2512317,2512316,9602111,9602112,1612303,3532407 3532408,1542257,2582375,9532060,9532059,9512049,8582017,9522056,1522237,1512222 1512223,0572187,2542331,2542335,2542336,0552170,0552171,9622125,9622136,9622137 1542256,2522321,3522484,0522149,9562172,1622310,1622311,1522241,7627001,0542163 2532329,8527004,9582093,2522325,8622036,1572283,8537009,8537010,1512224,1512225 2552342,2562353,2562352,8592027,9602110,9582100,2572369,9522052,1552261,9522054 9522053,1532252,8572012,8572011,8527005,8527006,8527003,8612032,9532062,2542340 2512312,0592201,9582094,1532251,0592200,0612215,9592108,8622043,9512044,1592289 1512226,0602209,9612117,9522055,9572087,0532155,8592026,9552077,1522238,8622042 1622307,1512231,1592288,0582197,1572271,2572388,9532066,1592285,0572188,1512232 1532248,1532249,0522151,8582018,8572013,8592023,2542337,3622137,3622138,4522195 3592096,3532035,3532036,5612601,5592557 数量 :475台 出荷時期 :平成8年 10月〜平成17年 2月
製造販売業者等名称
製造販売業者の名称 :東芝メディカルシステムズ株式会社 製造販売業者の住所 :栃木県大田原市下石上1385番地 薬事の業態 :第一種医療機器製造販売業 製造販売業許可番号 :09B1X00003 製造所の名称 :東芝メディカルシステムズ株式会社 製造所の住所 :栃木県大田原市下石上1385番地 薬事の業態 :医療機器製造業 製造業許可番号 :09BZ006008
改修理由
X線循環器システムにて使用しておりますX線用TV装置 MTV-500A において、カメラヘッドからの 映像信号を画像処理している基板上のICの未使用ゲートの入力端子の処理方法が不適切であることが 判明しました。 このため、IC未使用ゲートの動作安定化のため正しい使用方法に変更する対策を改修として実施させ ていただきます。
危惧される具体的な健康被害
X線用TV装置の映像信号を処理している基板上のICの未使用ゲートの不適切な使用により、X線用 TV装置の動作に影響を与えることはないと考えております。 そのため、重篤な健康被害が発生することはないと考えております。
改修開始年月日
平成18年 9月 8日 (情報提供の開始)
効能・効果又は用途等
X線蛍光増倍管(I.I.)の出力面に表示された画像を撮像し、ディジタル画像信号を出力する装置で、 医師もしくは放射線技師等有資格者が使用します。
その他
納入しました医療機関等は全て特定しており、全対象装置に改修を実施します。